-----BEGIN PRIVACY-ENHANCED MESSAGE----- Proc-Type: 2001,MIC-CLEAR Originator-Name: webmaster@www.sec.gov Originator-Key-Asymmetric: MFgwCgYEVQgBAQICAf8DSgAwRwJAW2sNKK9AVtBzYZmr6aGjlWyK3XmZv3dTINen TWSM7vrzLADbmYQaionwg5sDW3P6oaM5D3tdezXMm7z1T+B+twIDAQAB MIC-Info: RSA-MD5,RSA, R9T8158rZxlcD6hpxFnpJWkPUQ+2yHlNtYWPAmVAyRmTVJRcTPV84z8HCZcmF8EV YHE0jU1KG235deSKWYpIDQ== 0001299933-07-003872.txt : 20070627 0001299933-07-003872.hdr.sgml : 20070627 20070627172536 ACCESSION NUMBER: 0001299933-07-003872 CONFORMED SUBMISSION TYPE: 8-K PUBLIC DOCUMENT COUNT: 2 CONFORMED PERIOD OF REPORT: 20070626 ITEM INFORMATION: Regulation FD Disclosure FILED AS OF DATE: 20070627 DATE AS OF CHANGE: 20070627 FILER: COMPANY DATA: COMPANY CONFORMED NAME: STAAR SURGICAL CO CENTRAL INDEX KEY: 0000718937 STANDARD INDUSTRIAL CLASSIFICATION: OPHTHALMIC GOODS [3851] IRS NUMBER: 953797439 STATE OF INCORPORATION: DE FISCAL YEAR END: 0101 FILING VALUES: FORM TYPE: 8-K SEC ACT: 1934 Act SEC FILE NUMBER: 000-11634 FILM NUMBER: 07944515 BUSINESS ADDRESS: STREET 1: 1911 WALKER AVE CITY: MONROVIA STATE: CA ZIP: 91016 BUSINESS PHONE: 6263037902 MAIL ADDRESS: STREET 1: 1911 WALKER AVE CITY: MONROVIA STATE: CA ZIP: 91016 FORMER COMPANY: FORMER CONFORMED NAME: STAAR SURGICAL COMPANY DATE OF NAME CHANGE: 19920703 8-K 1 htm_21135.htm LIVE FILING STAAR Surgical Company (Form: 8-K)  

 


UNITED STATES
SECURITIES AND EXCHANGE COMMISSION

WASHINGTON, D.C. 20549

FORM 8-K

CURRENT REPORT

Pursuant to Section 13 or 15(d) of the Securities Exchange Act of 1934

     
Date of Report (Date of Earliest Event Reported):   June 26, 2007

STAAR Surgical Company
__________________________________________
(Exact name of registrant as specified in its charter)

     
Delaware 0-11634 95-3797439
_____________________
(State or other jurisdiction
_____________
(Commission
______________
(I.R.S. Employer
of incorporation) File Number) Identification No.)
      
1911 Walker Ave, Monrovia, California   91016
_________________________________
(Address of principal executive offices)
  ___________
(Zip Code)
     
Registrant’s telephone number, including area code:   626-303-7902

Not Applicable
______________________________________________
Former name or former address, if changed since last report

 

Check the appropriate box below if the Form 8-K filing is intended to simultaneously satisfy the filing obligation of the registrant under any of the following provisions:

[  ]  Written communications pursuant to Rule 425 under the Securities Act (17 CFR 230.425)
[  ]  Soliciting material pursuant to Rule 14a-12 under the Exchange Act (17 CFR 240.14a-12)
[  ]  Pre-commencement communications pursuant to Rule 14d-2(b) under the Exchange Act (17 CFR 240.14d-2(b))
[  ]  Pre-commencement communications pursuant to Rule 13e-4(c) under the Exchange Act (17 CFR 240.13e-4(c))


Item 7.01 Regulation FD Disclosure.

Item 8.01 Other Events
On June 26, 2007, the Company received a Warning Letter from the U.S. Food and Drug Administration ("FDA") citing four areas of noncompliance noted during an inspection of the Company's clinical study procedures, practices, and documentation related to the Toric Implantable Collamer Lens ("TICL"). The inspection was conducted by the Bioresearch Monitoring Program of the FDA Office of Regulatory Affairs ("BIMO") between February 15 and March 14, 2007. The Warning Letter notes deviations from FDA regulations that occurred between 2002 and 2005, which were among eight matters observed in the Inspectional Observations on FDA Form 483 received by the Company at the conclusion of the BIMO inspection. The Company described these observations in its Current Report on Form 8-K filed with the SEC on March 14, 2007, and provided the FDA with a written response to the Inspectional Observations on April 5, 2007.

Noting that the FDA staff found some of the Company’s responses in the April 5 letter inadequate, the Warning Letter has instructed the Company to provide additional corrective and preventative action plans and further information on some of the corrective actions described in the Company’s April 5 letter. The Company expects to provide its written response on or before July 18, 2007, the deadline for response stated in the Warning Letter.

David Bailey, President and CEO of STAAR Surgical Company, is expected to discuss the Warning Letter during a webcast presentation at the Jefferies Healthcare Conference, scheduled for 1:00 pm Eastern Time on Thursday, June 18, 2007. The webcast may be publicly accessed through the Investor Information page of the Company’s website at www.staar.com.

The Company has revised key processes involved in initiating and monitoring clinical studies, and anticipates that an enhancement of its processes and procedures, a detailed discussion of their intended corrective effect, and a satisfactory projected completi on date will adequately address the concerns of the FDA expressed in the Warning Letter. However, if the FDA does not find the Company’s response adequate, further administrative action could follow, including actions that could delay approval of the TICL or restrict the Corporation as a sponsor of clinical investigations.

BIMO inspections are part of a program designed to ensure that data and information contained in requests for Investigational Device Exemptions (IDE), Premarket Approval (PMA) applications, and Premarket Notification submissions (510k) are scientifically valid and accurate. Another objective of the program is to ensure that human subjects are protected from undue hazard or risk during the course of scientific investigations. While the past procedural violations noted in the Warning Letter are serious in nature and require comprehensive corrective and preventative actions, the Company does not believe that these nonconformities undermine the scientific validity and accurac y of its clinical data, or that human subjects were subjected to undue hazard or risk.

The description of the Warning Letter in this Report is qualified in its entirety by reference to the text of the Warning Letter, a copy of which is attached to this Report as Exhibit 99.1 and is incorporated herein by this reference.
All statements in this report that are not statements of historical fact are forward-looking statements, including statements about actions that may or may not be taken by the FDA, statements of the Company’s expectations or belief and any statements of assumptions underlying the foregoing. These statements are based on expectations and assumptions as of the date of this press release and are subject to numerous risks and uncertainties, which could cause actual results to differ materially from those described in the forward-looking statements. The risks and uncertainties include the FDA’s discretion in determining the adequacy of the Company’s response to the Warning Letter, the FDA’s discretion to issue additional warning letters or take other regulatory action as a result of inspectional observations or other data received by the agency, the risk that for serious violations of regulations related to clinical procedures FDA enforcement actions may result in penalties, injunctions or other measures, and the general risks to the Company’s business that result from FDA regulation, which are described in greater detail in our Quarterly Report on Form 10-Q filed on May 9, 2007, under the headings "Risk Factors - We are subject to extensive government regulation, which increases our costs and could prevent us from selling our products," and "Risk Factors - FDA compliance issues have harmed our reputation, and we expect to devote significant resources to maintaining compliance in the future," and the other risks describe in the Risk Factors section of that report and detailed from time to time in our other reports filed with the Securities and Exchang e Commission. STAAR Surgical Company assumes no obligation to update these forward-looking statements to reflect future events or actual outcomes and does not intend to do so.






SIGNATURES

Pursuant to the requirements of the Securities Exchange Act of 1934, the registrant has duly caused this report to be signed on its behalf by the undersigned hereunto duly authorized.

         
    STAAR Surgical Company
          
June 27, 2007   By:   /s/David Bailey
       
        Name: David Bailey
        Title: President and Chief Executive Officer


Exhibit Index


     
Exhibit No.   Description

 
99.1
  Warning Letter dated June 26, 2007 from U.S. Food and Drug Administration.
EX-99.1 2 exhibit1.htm EX-99.1 EX-99.1

*** Indicates information redacted from public disclosure pursuant to the

policies of the FDA.

DEPARTMENT OF HEALTH & HUMAN SERVICES            Public Health Service

     

Food and Drug Administration
9200 Corporate Blvd.
Rockville MD 20850

JUN 26 2007

WARNING LETTER

Via Federal Express

David Bailey, President
Staar Surgical Company
1911 Walker Avenue
Monrovia, CA 91016-4846

Dear Mr. Bailey:

This Warning Letter is to inform you of objectionable conditions observed during the Food and Drug Administration (FDA) inspection conducted at your firm from February 15 through March 14, 2007, by an investigator from the FDA Los Angeles District Office. The purpose of this inspection was to determine whether activities and procedures related to your participation in the clinical study titled Clinical Study Protocol — STAAR Surgical Implantable Toric Phakic ICL (TICL) for Myopic/Astigmatic Patients under IDE # G010252, and in support of PMA P030016/S001, complied with applicable federal regulations. The ICL (implantable contact lens) used for the study is a device as that term is defined in section 201(h) of the Federal Food, Drug and Cosmetic Act (the Act), 21 U.S.C. 321(h). This letter also requests prompt corrective action to address the violations cited and discusses your April 5, 2007, written response to the noted violations.

The inspection was conducted under a program designed to ensure that data and information contained in requests for Investigational Device Exemptions (IDE), Premarket Approval (PMA) applications, and Premarket Notification submissions (510(k)) are scientifically valid and accurate. Another objective of the program is to ensure that human subjects are protected from undue hazard or risk during the course of scientific investigations.

Our review of the inspection report prepared by the district office revealed serious violations of Title 21, Code of Federal Regulations (21 CFR) Part 812 — Investigational Device Exemptions, Part 50 — Protection of Human Subjects, and Section 520(g) (21 U.S.C. 360j(g)) of the Act. At the close of the inspection, the FDA investigator presented an inspectional observations form FDA 483 for your review and discussed the observations listed on the form with you. The deviations noted on the FDA 483, your written response, and our subsequent review of the inspection report are discussed below:

1

Page 2 — David Bailey

1.   Failure to submit an IDE application to FDA, ensure IRB review and approval are obtained, and ensure FDA approval of an IDE application prior to beginning an investigation [21 CPR 812.40 & 21 CFR 812.42].

You failed to adhere to the above noted regulations. Specifically:

      a.) Staar Surgical sponsored and initiated a clinical research study with a significant risk device without submitting an IDE application to FDA. A study site in ***, which was listed as a participant in the TICL study being conducted under IDE G010252, conducted its study under a different protocol that was designed to achieve a different objective than the study approved under IDE G010252. The study protocol under IDE, G010252, as approved by the FDA, was for a non-randomized, open-label study to assess the effectiveness of the device. The study at the site in ***, which listed *** as the clinical investigator, was for a randomized study in which subjects were randomly assigned to receive either the TICL implant or PRK (photorefractive keratectomy) with Mitomycin C. The study objectives were to assess and compare clinical outcomes of subjects receiving the two treatments.

*** claimed on his IRB submission that this study was “part of an IDE study sponsored by and with the consent of Staar Surgical” and claimed that it was being conducted under IDE G010252. Staar Surgical provided the devices for the study, which was approved by the IRB on 9/17/03. *** subsequently implanted the TICL into 43 eyes. The data for these procedures have been included in your PMA submission P030016/S001 to the FDA.

In your response letter, dated April 5, 2007, you stated that “this issue was identified by the STAAR clinical team during their visit February 16/17, 2006,” and that you are “in active discussion with our site monitor contractor to establish a very specific schedule of activities to be undertaken.” This response is not adequate, in that it does not address any corrective and preventive actions regarding initiation of a significant risk study without an IDE. Please provide documentation of a corrective action plan, specific to this issue.

      b.) At least 4 of the 7 clinical study sites did not obtain IRB approval for the correct version of the study protocol. ***, ***, ***, and *** used *** as their IRB of record. *** approved the protocol .version dated 12/3/01 for *** on 4/5/02, and for each of the other three study sites on 2/8/02. The FDA approved a revised version of the protocol on 1/3/02, and Staar Surgical finalized the revisions in the protocol version dated 2/20/02. ***, the Staar Surgical Vice President of Regulatory Affairs and Quality Assurance, provided a document to the FDA investigator that stated the study was “activated” under the February 2002 version of the protocol, (with the FDA-requested revisions). ***, Senior Clinical Research Associate, who was present during the inspection, also told the FDA investigator that *** confirmed with him that they never reviewed or approved the 2/20/02 version of the protocol for these 4 study sites. Each of

Page 3 — David Bailey

these 4 study sites enrolled several subjects under the 12/3/01 (non-FDA approved) version of the protocol.

In your response letter, you stated that a new procedure is being developed to cover annual renewals and supplemental approvals of revisions to protocols. Please provide us with a copy of this procedure once it is finalized, and please include an explanation and supporting documentation regarding training of appropriate staff on the new procedure.

2.   Failure to secure the investigators’ compliance with the investigational plan and applicable FDA regulations [21 CFR 812.46(a)].

Staar Surgical failed to ensure that all clinical investigators participating in the study adhered to the investigational plan and FDA regulations. Examples of this failure include:

      a.) The Patient Eligibility Checklist Case Report Form (CRF) provided to *** clinical site in *** listed a requirement for “direct and retro illumination photograph.” The CRF also stated that “all answers must be YES to proceed” and “if the patient does not satisfy ALL the eligibility criteria above DO NOT PROCEED FURTHER.” On all 43 CRFs for the subjects’ eyes implanted by ***, the question for “direct and retro illumination photographs” was marked “N/A.” *** told the FDA investigator that this study site had been given verbal approval to proceed without these photographs, and that the site did not have the necessary equipment to take these photographs.

Furthermore, in the Clindex Database maintained for this study, the responses that were marked “N/A” on the CRFs were entered as “1” (indicating a YES answer) or left blank in the data listing. The original validation for the database required that an entry marked as “2” (no photographs) or left blank would result in a message that “this patient might become disqualified.” The database coding was changed on 10/27/04 to show no message when any entry was made, and to default to a “1” no matter what entry was made.

In your response, you noted that direct and retro-illumination photographs were not required in the study protocol, and the eligibility checklist CRF was inadvertently carried over from the previous study. FDA regulations consider an investigational plan to include the written protocol [21 CFR 812.25(b)]. The CRFs are considered part of the investigational plan. If the requirement for direct and retro-illumination photographs was not a requirement for the study, it should not have been included on the CRF and in the study database. Your response also stated that your IT department is developing a new procedure covering changes to validated databases. This is not an adequate response in that it does not address any corrective or preventative actions for the above stated violation. Please provide an explanation and supporting documentation regarding how Staar Surgical will ensure that all study documents correspond to the approved protocol, and how study site non-compliance issues will be addressed for future studies.

2

Page 4 — David Bailey

      b.) A study site under *** in ***used a non-IRB approved consent form for the pre-operative YAG Laser Iridotomy portion of the study, for 24 of the 57 subjects enrolled in the study at that site. The IRB had approved the consent form on 2/8/02. This violation occurred during the period of 9/17/02 through 12/30/05, which is after the IRB approved the consent form. Furthermore, Staar Surgical did not identify this violation until a monitoring visit conducted in August 2006, nearly four years after the date of the first violation.

In your response, you stated that “appropriate action was swiftly implemented to remediate the situation” following a site visit in August 2006. You also stated that you are “amending arrangements with our site monitoring contractor to avoid recurrence of similar omissions.” This response is not adequate in that it does not address any corrective or preventative actions for the violation. Please provide an explanation and supporting documentation regarding how Staar Surgical will ensure that study sites are using the correct IRB-approved consent forms for all enrolled subjects.

3.   Failure to immediately conduct an evaluation of all unexpected adverse device effects [21 CFR 812.46(b)].

According to Staar Surgical’s procedure titled “Clinical Trials Adverse Event Reporting,” effective 8/30/2005, any unexpected adverse device effects are to be reported to the FDA within 10 working days of initial receipt of the report. Anticipated adverse events will be reported annually in the IDE annual report. The SOP also notes that a change in frequency or severity of anticipated adverse events will be evaluated by the Medical Reviewer to determine whether it becomes an unanticipated adverse event. At least three adverse device effects were reported during the study, which were not immediately investigated to determine if they were expected or unexpected. Specifically:

      a.) Subject 00142 experienced subcapsular cataract in the left eye, as reported to Staar Surgical on 1/8/04. Subject 00103 experienced retinal detachment requiring surgical treatment on 7/8/04. There was no documentation that these events were evaluated by your Medical Reviewer to determine if they met the 10-day reportability requirement.

      b.) Subject 00021 experienced slight Iris Bombay, enlarged pupil, and darkened iris on 1/31/03. This event was not evaluated by the Medical Reviewer until 1/28/04, which is nearly a year after the event occurred.

In addition, during the FDA inspection, your Medical Reviewer, *** reported that he discarded all documentation regarding his evaluation of adverse events. Your Adverse Event procedure also states that “the Clinical Department will trend anticipated adverse events, as appropriate. If there is an increase in the frequency or severity of an event, the information will be forwarded to the Medical Reviewer.” *** told *** during the FDA inspection that he did not know how Staar Surgical established the expected frequency and severity of anticipated adverse events, and *** could not provide documentation that trending of adverse events occurred.

3

Page 5 — David Bailey

In your response, you stated that your adverse event reporting procedure “is being revised to enhance controls along the lines discussed.” Please provide us with a copy of this procedure once it is finalized, and please include an explanation and supporting documentation regarding training of appropriate staff on the new procedure.

4.   Failure to submit required reports and information to the FDA for an investigation being conducted under an IDE, including reports of prior investigations. [21 CFR 812.27]

Staar Surgical failed to adhere to the above noted regulations. Specifically, during the FDA inspection, *** told the FDA investigator that a similar study had been conducted with 5 subjects in Europe prior to conduct of this study. The protocol was titled “Investigational Plan Pilot Study ICL™ Toric™ IDE # 012 AG, The Implantable Toric Contact Lens for Patients with Myopia and Astigmatism, version May 11, 1999.” Information regarding this study was not provided to FDA in the IDE submission, and *** could not provide any information regarding the outcome of this study.

In your response, you stated that you have asked the Staar Surgical team in the Swiss facility to notify you of any records related to the conduct of the pilot study in Europe. Please provide us with this information once it is received.

In addition to the violation above, you notified FDA of an IDE change on October 8, 2002, in which you added *** as one of the clinical investigators for the IDE study. In this notification letter, you stated, “No other changes are being made to the Toric Implantable Contact Lens (TICL) Clinical Investigational Plan approved by the agency on January 3, 2002.” This statement appears to be false since, as previously mentioned, the study conducted by *** differed significantly from the study approved under the IDE. *** provided a memo to Staar Surgical, dated February 17, 2006, in which he stated “early in the discussion of plans to include NMCSD as a clinical site, we decided we wanted to expand our participation to include randomization of study participants... We had discussion with ***, by teleconference...we explained our plan to include randomization to procedure and *** provided enthusiastic agreement.”

The violations described above are not intended to be an all-inclusive list of problems that may exist with your clinical study. It is your responsibility as a study sponsor to ensure compliance with the Act and applicable regulations.

Within fifteen (15) working days of receiving this letter, please provide written documentation of the additional actions you have taken or will take to correct these violations and prevent the recurrence of similar violations in current or future studies for which you are the study sponsor. Any submitted corrective action plan must include projected completion dates for each action to be accomplished. Failure to respond to this letter and take appropriate corrective action could result in the FDA taking regulatory action without further notice to you. Please send your response to:

4

Page 6 — David Bailey

Food and Drug Administration
Center for Devices and Radiological Health
Office of Compliance, Division of Bioresearch Monitoring, HFZ-311
9200 Corporate Boulevard, Rockville, Maryland 20850
Attention: ***, MSN, Chief, Special Investigations Branch.

A copy of this letter has been sent to FDA’s Los Angeles District Office, 19701 Fairchild, Irvine, CA 92612. We request that a copy of your response also be sent to that office.

If you have any questions, please contact *** at*** or at ***.

Sincerely yours,

Office of Compliance
Center for Devices and

Radiological Health

5 -----END PRIVACY-ENHANCED MESSAGE-----